AQ2013-06 批次 TSH 檢驗院際品管結果報告 |
送檢批號:AQ2013-06 調查期間:2013 年 06 月 03 ~
10 日 參加單位數:21
檢驗單位 代碼 |
確診醫院 代碼 |
檢體方 法編碼 |
檢體試 劑編碼 |
Sample
S1 |
Sample
S2 |
||||||
(mIU/L) |
D 1 |
Z 2 |
Da% 3 |
(mIU/L) |
D 1 |
Z 2 |
Da% 3 |
||||
RH01b |
RH01 |
1 |
3 |
7.9 |
-1.4 |
-1.9 |
-63% |
1.6 |
0.4 |
4.0 |
139% |
RH07b |
RH07 |
1 |
3 |
7.9 |
-1.4 |
-1.9 |
-63% |
1.5 |
0.3 |
3.0 |
104% |
RH14 |
RH14 |
1 |
3 |
7.5 |
-1.8 |
-2.4 |
-81% |
1.4 |
0.2 |
2.0 |
69% |
RH15 |
RH15 |
1 |
3 |
7.7 |
-1.6 |
-2.2 |
-72% |
1.5 |
0.3 |
3.0 |
104% |
CL009 |
CL009 |
1 |
3 |
7.7 |
-1.6 |
-2.2 |
-72% |
1.4 |
0.2 |
2.0 |
69% |
CL010 |
CL010 |
1 |
3 |
7.5 |
-1.8 |
-2.4 |
-81% |
1.4 |
0.2 |
2.0 |
69% |
CL012 |
CL012 |
1 |
3 |
8.0 |
-1.3 |
-1.8 |
-58% |
1.7 |
0.5 |
5.0 |
174% |
CL013 |
CL013 |
1 |
5 |
9.3 |
0.0 |
0.0 |
0% |
1.0 |
-0.2 |
-2.0 |
-69% |
CL014a |
CL014 |
1 |
3 |
7.2 |
-2.1 |
-2.8 |
-94% |
1.4 |
0.2 |
2.0 |
69% |
CL015 |
CL015 |
1 |
3 |
8.6 |
-0.7 |
-0.9 |
-31% |
1.8 |
0.6 |
6.0 |
208% |
Median of RIA Group |
7.8 |
1.5 |
|||||||||
Range of RIA Group |
7.2 – 9.3 |
1.0 – 1.8 |
|||||||||
Robust average4 of RIA Group |
7.9 ( n = 10 ) |
1.5 ( n = 10 ) |
|||||||||
Robust SD4 of RIA Group |
0.53 |
0.19 |
|||||||||
CV of RIA Group |
6.7% |
12.7% |
|||||||||
Median of European RIA labs |
9.5 ( n = 107 ) |
1.2 ( n = 102 ) |
|||||||||
Range of European labs (16% – 84%) |
9.1 – 10.0 |
1.1 – 1.5 |
RH01a |
RH01 |
3 |
9 |
8.3 |
-1.0 |
-1.4 |
-45% |
1.0 |
-0.2 |
-2.0 |
-69% |
RH02c |
RH02 |
3 |
1 |
10.4 |
1.1 |
1.5 |
49% |
0.9 |
-0.3 |
-3.0 |
-104% |
RH06 |
RH06 |
3 |
9 |
8.2 |
-1.1 |
-1.5 |
-49% |
0.9 |
-0.3 |
-3.0 |
-104% |
RH07a |
RH07 |
3 |
2 |
8.3 |
-1.0 |
-1.4 |
-45% |
1.0 |
-0.2 |
-2.0 |
-69% |
RH12 |
RH12 |
3 |
2 |
10.4 |
1.1 |
1.5 |
49% |
1.1 |
-0.1 |
-1.0 |
-35% |
RH19 |
RH19 |
3 |
2 |
10.9 |
1.6 |
2.2 |
72% |
1.1 |
-0.1 |
-1.0 |
-35% |
RH20 |
RH20 |
3 |
7 |
9.1 |
-0.2 |
-0.3 |
-9% |
1.1 |
-0.1 |
-1.0 |
-35% |
CL005 |
CL005 |
3 |
9 |
8.3 |
-1.0 |
-1.4 |
-45% |
1.0 |
-0.2 |
-2.0 |
-69% |
CL006a |
CL006 |
3 |
7 |
9.0 |
-0.3 |
-0.4 |
-13% |
1.1 |
-0.1 |
-1.0 |
-35% |
CL008 |
CL008 |
3 |
1 |
9.6 |
0.3 |
0.4 |
13% |
0.8 |
-0.4 |
-4.0 |
-139% |
CL011 |
CL011 |
3 |
13 |
9.4 |
0.1 |
0.1 |
4% |
1.3 |
0.1 |
1.0 |
35% |
Median of CLIA Group |
9.1 |
1.0 |
|||||||||
Range of CLIA Group |
8.2 – 10.9 |
0.8 – 1.3 |
|||||||||
Robust average4 of CLIA Group |
9.3 ( n = 11 ) |
1.0 ( n = 11 ) |
|||||||||
Robust SD4 of CLIA Group |
1.10 |
0.13 |
|||||||||
CV of CLIA Group |
11.8% |
13.0% |
|||||||||
Median of European labs |
9.1 ( n = 1180 ) |
1.2 ( n = 1145 ) |
|||||||||
Range of European CLIA labs ( 16% – 84% ) |
8.5 – 10.3 |
1.0 – 1.3 |
Median of All Labs |
8.3 |
1.1 |
Range of All Labs |
7.2 – 10.9 |
0.8 – 1.8 |
Robust average4 of All Labs |
8.6 ( n = 21 ) |
1.2 ( n = 21) |
Robust SD4 of All Labs |
1.07 |
0.31 |
CV of All Labs |
12.3% |
25.8% |
Median of All European Labs |
9.3 |
1.2 |
Range of All European Labs (16% – 84%) |
8.4 – 10.4 |
1.0 – 1.3 |
Mean of All European Labs |
9.3 ( n = 1411 ) |
1.2 ( n = 1377 ) |
SD of All European Labs |
1.31 |
0.17 |
CV of All European Labs |
14.1% |
14.4% |
1. D = X - Xa 2. Zscore = D / Target SD 3. Da% (Your result from
target as percentages of allowed deviation) = D/(Xa
x MAD) x 100% 4. Robust results were
calculated by Algorithm A according to ISO 13528:2005 and IUPAC/CITAC Guide:
Selection and use of proficiency testing scheme for a limited number of
participants. Pure Appl Chem
2010;82:1099-135. 5. The Target value (Xa)
is the median of the results of this control material used
in European surveys. 6. Maximum allowable deviation
(MAD) 24%, Target SD = 8% x Xa |
|
Sample
S1 |
Sample
S2 |
Target value (Xa) 5 (
mIU/L ) |
9.3 ( n = 1411 ) |
1.2 ( n = 1377 ) |
Target SD 6 |
0.74 |
0.10 |
Method Name |
Method Code |
Laboratory |
Radioimmunoassay ( Sandwich - type ) |
1 |
RH01b, RH07b, RH14, RH15, CL009, CL010, CL012,
CL013, CL014a, CL015, |
Chemiluminescent Immunoassay ( Sandwich - type ) |
3 |
RH01a, RH02c, RH06, RH07a, RH12, RH19, RH20,
CL005, CL006a, CL008, CL011, |
Reagent Code |
Laboratory |
|
Abbott
Architect |
1 |
CL008,
RH02c, |
Beckman
Coulter Access |
2 |
RH07a,
RH12, RH19, |
Cisbio
RIA-gnost |
3 |
CL009,
CL010, CL012, CL014a, CL015, RH01b, RH07b, RH14, RH15, |
Immunotech |
5 |
CL013,
|
Siemens
ADVIA Centaur |
7 |
CL006a,
RH20, |
Siemens
Immulite 2000 |
9 |
CL005,
RH01a, RH06, |
Roche |
13 |
CL011,
|